On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysis
The aim of this study was to analyse the reproducibility of off-axis integrated cavity output spectroscopy (OA-ICOS)-derived δ<sup>2</sup>H and δ<sup>18</sup>O measurements on a set of 35 water samples by comparing the performance of four laser spectroscopes with the performa...
Main Authors: | , , , , , , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Copernicus Publications
2010-08-01
|
Series: | Hydrology and Earth System Sciences |
Online Access: | http://www.hydrol-earth-syst-sci.net/14/1551/2010/hess-14-1551-2010.pdf |
id |
doaj-f635b9b9315a46ba8dc2fa20a5317f58 |
---|---|
record_format |
Article |
spelling |
doaj-f635b9b9315a46ba8dc2fa20a5317f582020-11-24T23:09:03ZengCopernicus PublicationsHydrology and Earth System Sciences1027-56061607-79382010-08-011481551156610.5194/hess-14-1551-2010On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysisM. BonazzaB. M. C. FischerM. BorgaA. GobbiT. A. BogaardS. WredeM. ŠandaB. StenniD. PennaZ. ChárováThe aim of this study was to analyse the reproducibility of off-axis integrated cavity output spectroscopy (OA-ICOS)-derived δ<sup>2</sup>H and δ<sup>18</sup>O measurements on a set of 35 water samples by comparing the performance of four laser spectroscopes with the performance of a conventional mass spectrometer under typical laboratory conditions. All samples were analysed using three different schemes of standard/sample combinations and related data processing to assess the improvement of results compared with mass spectrometry. The repeatability of the four OA-ICOS instruments was further investigated by multiple analyses of a sample subset to evaluate the stability of δ<sup>2</sup>H and δ<sup>18</sup>O measurements. <br></br> Results demonstrated an overall agreement between OA-ICOS-based and mass spectrometry-based measurements for the entire dataset. However, a certain degree of variability existed in precision and accuracy between the four instruments. There was no evident bias or systematic deviations from the mass spectrometer values, but random errors, which were apparently not related to external factors, significantly affected the final results. Our investigation revealed that analytical precision ranged ±from ±0.56‰ to ±1.80‰ for δ<sup>2</sup>H and from ±0.10‰ to ±0.27‰ for δ<sup>18</sup>O measurements, with a marked variability among the four instruments. The overall capability of laser instruments to reproduce stable results with repeated measurements of the same sample was acceptable, and there were general differences within the range of the analytical precision for each spectroscope. Hence, averaging the measurements of three identical samples led to a higher degree of accuracy and eliminated the potential for random deviations. http://www.hydrol-earth-syst-sci.net/14/1551/2010/hess-14-1551-2010.pdf |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
M. Bonazza B. M. C. Fischer M. Borga A. Gobbi T. A. Bogaard S. Wrede M. Šanda B. Stenni D. Penna Z. Chárová |
spellingShingle |
M. Bonazza B. M. C. Fischer M. Borga A. Gobbi T. A. Bogaard S. Wrede M. Šanda B. Stenni D. Penna Z. Chárová On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysis Hydrology and Earth System Sciences |
author_facet |
M. Bonazza B. M. C. Fischer M. Borga A. Gobbi T. A. Bogaard S. Wrede M. Šanda B. Stenni D. Penna Z. Chárová |
author_sort |
M. Bonazza |
title |
On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysis |
title_short |
On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysis |
title_full |
On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysis |
title_fullStr |
On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysis |
title_full_unstemmed |
On the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>H and δ<sup>18</sup>O isotopic analysis |
title_sort |
on the reproducibility and repeatability of laser absorption spectroscopy measurements for δ<sup>2</sup>h and δ<sup>18</sup>o isotopic analysis |
publisher |
Copernicus Publications |
series |
Hydrology and Earth System Sciences |
issn |
1027-5606 1607-7938 |
publishDate |
2010-08-01 |
description |
The aim of this study was to analyse the reproducibility of off-axis integrated cavity output spectroscopy (OA-ICOS)-derived δ<sup>2</sup>H and δ<sup>18</sup>O measurements on a set of 35 water samples by comparing the performance of four laser spectroscopes with the performance of a conventional mass spectrometer under typical laboratory conditions. All samples were analysed using three different schemes of standard/sample combinations and related data processing to assess the improvement of results compared with mass spectrometry. The repeatability of the four OA-ICOS instruments was further investigated by multiple analyses of a sample subset to evaluate the stability of δ<sup>2</sup>H and δ<sup>18</sup>O measurements. <br></br> Results demonstrated an overall agreement between OA-ICOS-based and mass spectrometry-based measurements for the entire dataset. However, a certain degree of variability existed in precision and accuracy between the four instruments. There was no evident bias or systematic deviations from the mass spectrometer values, but random errors, which were apparently not related to external factors, significantly affected the final results. Our investigation revealed that analytical precision ranged ±from ±0.56‰ to ±1.80‰ for δ<sup>2</sup>H and from ±0.10‰ to ±0.27‰ for δ<sup>18</sup>O measurements, with a marked variability among the four instruments. The overall capability of laser instruments to reproduce stable results with repeated measurements of the same sample was acceptable, and there were general differences within the range of the analytical precision for each spectroscope. Hence, averaging the measurements of three identical samples led to a higher degree of accuracy and eliminated the potential for random deviations. |
url |
http://www.hydrol-earth-syst-sci.net/14/1551/2010/hess-14-1551-2010.pdf |
work_keys_str_mv |
AT mbonazza onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT bmcfischer onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT mborga onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT agobbi onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT tabogaard onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT swrede onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT msanda onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT bstenni onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT dpenna onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis AT zcharova onthereproducibilityandrepeatabilityoflaserabsorptionspectroscopymeasurementsfordsup2suphanddsup18supoisotopicanalysis |
_version_ |
1725611807597395968 |