In vitro comparison of working length determination using three different electronic apex locators

Background: The aim of this study was to compare the accuracy of the apex-locating functions of DentaPort ZX, Raypex 5 and Endo Master electronic apex locators (EALs) in vitro. Materials and Methods: Thirty extracted human single-rooted teeth with mature apices were used for the study. The real wo...

Full description

Bibliographic Details
Main Authors: Alper Kuştarci, Dilara Arslan, Demet Altunbaş
Format: Article
Language:English
Published: Wolters Kluwer Medknow Publications 2014-01-01
Series:Dental Research Journal
Subjects:
Online Access:http://www.drjjournal.net/article.asp?issn=1735-3327;year=2014;volume=11;issue=5;spage=568;epage=573;aulast=Kuştarci
Description
Summary:Background: The aim of this study was to compare the accuracy of the apex-locating functions of DentaPort ZX, Raypex 5 and Endo Master electronic apex locators (EALs) in vitro. Materials and Methods: Thirty extracted human single-rooted teeth with mature apices were used for the study. The real working length (RWL) was established by subtracting 0.5 mm from the actual root canal length. All teeth were mounted in an alginate model that was especially developed to test the EALs and the teeth were then measured with each EAL. The results were compared with the corresponding RWL, which was subtracted from the electronically determined distance. Data were analyzed using a paired-samples t-test, a Chi-square test and a repeated measure analysis of variance evaluation at the 0.05 level of significance. Results: Statistical analysis showed that no significant difference was found among all EALs (P > 0.05). Conclusion: The accuracy of the EALs was evaluated and all of the devices showed an acceptable determination of electronic working length between the ranges of ±0.5 mm.
ISSN:1735-3327
2008-0255