Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometer
AIM: To analyze the accuracy of non-contact tonometer(NCT)and ICare rebound tonometer(RBT)in measuring intraocular pressure in glaucoma patients. <p>METHODS:A retrospective proportional control study. Totally 113 patients(185 eyes)with glaucoma were studied. Intraocular pressure was measured b...
Main Authors: | , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Press of International Journal of Ophthalmology (IJO PRESS)
2019-08-01
|
Series: | Guoji Yanke Zazhi |
Subjects: | |
Online Access: | http://ies.ijo.cn/cn_publish/2019/8/201908034.pdf |
id |
doaj-fbe2ef97307542e39fa51f82efe898bc |
---|---|
record_format |
Article |
spelling |
doaj-fbe2ef97307542e39fa51f82efe898bc2020-11-25T01:45:42ZengPress of International Journal of Ophthalmology (IJO PRESS)Guoji Yanke Zazhi1672-51231672-51232019-08-011981406141010.3980/j.issn.1672-5123.2019.8.34Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometerYu-Han Bao0Peng Lyu1Wen-Fang Zhang2Wan-Min Wu3The Second Clinical Medical College of Lanzhou University, Lanzhou 730000, Gansu Province, China; Department of Ophthalmology, Lanzhou University Second Hospital, Lanzhou 730030, Gansu Province, ChinaDepartment of Ophthalmology, Lanzhou University Second Hospital, Lanzhou 730030, Gansu Province, ChinaDepartment of Ophthalmology, Lanzhou University Second Hospital, Lanzhou 730030, Gansu Province, ChinaShanxi Eye Hospital, Taiyuan 030002, Shanxi Province, ChinaAIM: To analyze the accuracy of non-contact tonometer(NCT)and ICare rebound tonometer(RBT)in measuring intraocular pressure in glaucoma patients. <p>METHODS:A retrospective proportional control study. Totally 113 patients(185 eyes)with glaucoma were studied. Intraocular pressure was measured by RBT, NCT and Goldmann Applanation tonometer(GAT)at 9:00 and 16:00 and recorded as two sets of data, respectively. Their accuracy is evaluated by the rank sum test, and the international standard 8621 guide, Bland-Altman consistency analysis analyzed the consistency of the two groups of measurements, and the data were divided into two groups: abnormal intraocular pressure group(<10mmHg or >21mmHg)and normal intraocular pressure group(10-21mmHg). The consistency was analyzed between different groups. The spearman correlation analysis is adopted to analyze the correlation between ophthalmotonometers.<p>RESULTS: There were differences among the GAT, NCT and RBT measurements(<i>P</i><0.01). There was a strong correlation between the intraocular pressure measured value obtained by GAT and measured values obtained by RBT and NCT(<i>r<sub>s</sub></i>=0.71, 0.77, <i>P</i><0.001). The measured values of NCT and GAT are close to each other without statistically significant differences(<i>P</i>=0.92). Compared with GAT, RBT has a higher intraocular pressure measured value, and there are significant differences(<i>P</i><0.05). However, compared with GAT, both NCT and RBT have a wide range of limits of agreement, and they are(-6.2-6.0mmHg)and(-5.2-7.6mmHg)respectively. In normal intraocular pressure groups, the ranges of limits of agreement are(-5.9-5.9mmHg)and(-4.3-7.5mmHg)respectively. In abnormal intraocular pressure groups, the ranges of limits of agreement are(-7.3-6.4mmHg)and(-7.5-5.6mmHg)respectively. The evaluation carried out in accordance with ISO 8612 guides indicates that the abnormal values more than the 95% of consistency restriction of NCT and RBT in the three groups are(3.9%, 11.3%, 12.2%)and(26.3%, 11.3%, 12.2%)respectively.<p>CONCLUSION: Both NCT and RBT cannot simply replace the GAT to measure the intraocular pressure of glaucoma patients. In glaucoma patients, with the deviation of the measured value from the normal intraocular pressure range, the measurement error of NCT and RBT also increase.http://ies.ijo.cn/cn_publish/2019/8/201908034.pdfglaucomaintraocular pressurenon-contact tonometerICare rebound tonometerGoldmann applanation tonometry |
collection |
DOAJ |
language |
English |
format |
Article |
sources |
DOAJ |
author |
Yu-Han Bao Peng Lyu Wen-Fang Zhang Wan-Min Wu |
spellingShingle |
Yu-Han Bao Peng Lyu Wen-Fang Zhang Wan-Min Wu Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometer Guoji Yanke Zazhi glaucoma intraocular pressure non-contact tonometer ICare rebound tonometer Goldmann applanation tonometry |
author_facet |
Yu-Han Bao Peng Lyu Wen-Fang Zhang Wan-Min Wu |
author_sort |
Yu-Han Bao |
title |
Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometer |
title_short |
Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometer |
title_full |
Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometer |
title_fullStr |
Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometer |
title_full_unstemmed |
Accuracy of intraocular pressure measurements between non-contact tonometer and ICare rebound tonometer |
title_sort |
accuracy of intraocular pressure measurements between non-contact tonometer and icare rebound tonometer |
publisher |
Press of International Journal of Ophthalmology (IJO PRESS) |
series |
Guoji Yanke Zazhi |
issn |
1672-5123 1672-5123 |
publishDate |
2019-08-01 |
description |
AIM: To analyze the accuracy of non-contact tonometer(NCT)and ICare rebound tonometer(RBT)in measuring intraocular pressure in glaucoma patients. <p>METHODS:A retrospective proportional control study. Totally 113 patients(185 eyes)with glaucoma were studied. Intraocular pressure was measured by RBT, NCT and Goldmann Applanation tonometer(GAT)at 9:00 and 16:00 and recorded as two sets of data, respectively. Their accuracy is evaluated by the rank sum test, and the international standard 8621 guide, Bland-Altman consistency analysis analyzed the consistency of the two groups of measurements, and the data were divided into two groups: abnormal intraocular pressure group(<10mmHg or >21mmHg)and normal intraocular pressure group(10-21mmHg). The consistency was analyzed between different groups. The spearman correlation analysis is adopted to analyze the correlation between ophthalmotonometers.<p>RESULTS: There were differences among the GAT, NCT and RBT measurements(<i>P</i><0.01). There was a strong correlation between the intraocular pressure measured value obtained by GAT and measured values obtained by RBT and NCT(<i>r<sub>s</sub></i>=0.71, 0.77, <i>P</i><0.001). The measured values of NCT and GAT are close to each other without statistically significant differences(<i>P</i>=0.92). Compared with GAT, RBT has a higher intraocular pressure measured value, and there are significant differences(<i>P</i><0.05). However, compared with GAT, both NCT and RBT have a wide range of limits of agreement, and they are(-6.2-6.0mmHg)and(-5.2-7.6mmHg)respectively. In normal intraocular pressure groups, the ranges of limits of agreement are(-5.9-5.9mmHg)and(-4.3-7.5mmHg)respectively. In abnormal intraocular pressure groups, the ranges of limits of agreement are(-7.3-6.4mmHg)and(-7.5-5.6mmHg)respectively. The evaluation carried out in accordance with ISO 8612 guides indicates that the abnormal values more than the 95% of consistency restriction of NCT and RBT in the three groups are(3.9%, 11.3%, 12.2%)and(26.3%, 11.3%, 12.2%)respectively.<p>CONCLUSION: Both NCT and RBT cannot simply replace the GAT to measure the intraocular pressure of glaucoma patients. In glaucoma patients, with the deviation of the measured value from the normal intraocular pressure range, the measurement error of NCT and RBT also increase. |
topic |
glaucoma intraocular pressure non-contact tonometer ICare rebound tonometer Goldmann applanation tonometry |
url |
http://ies.ijo.cn/cn_publish/2019/8/201908034.pdf |
work_keys_str_mv |
AT yuhanbao accuracyofintraocularpressuremeasurementsbetweennoncontacttonometerandicarereboundtonometer AT penglyu accuracyofintraocularpressuremeasurementsbetweennoncontacttonometerandicarereboundtonometer AT wenfangzhang accuracyofintraocularpressuremeasurementsbetweennoncontacttonometerandicarereboundtonometer AT wanminwu accuracyofintraocularpressuremeasurementsbetweennoncontacttonometerandicarereboundtonometer |
_version_ |
1725023331833348096 |