Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study

碩士 === 中山醫學大學 === 生物醫學科學學系碩士班 === 100 === The traditional spiral scans of computer tomography using a single energy, and recently the Siemens CT using dual-energy scanning protocol apply to arterial phase scan of liver triple-phase examination, and improve the diagnosis rate of the lesion. The T...

Full description

Bibliographic Details
Main Authors: Po-Hung, 賴柏宏
Other Authors: 陳拓榮
Format: Others
Language:zh-TW
Published: 2012
Online Access:http://ndltd.ncl.edu.tw/handle/93808531008655773890
id ndltd-TW-100CSMU5114009
record_format oai_dc
spelling ndltd-TW-100CSMU51140092015-10-13T21:55:43Z http://ndltd.ncl.edu.tw/handle/93808531008655773890 Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study 探討雙源與320切電腦斷層肝三相之影像品質與輻射劑量:假體實驗 Po-Hung 賴柏宏 碩士 中山醫學大學 生物醫學科學學系碩士班 100 The traditional spiral scans of computer tomography using a single energy, and recently the Siemens CT using dual-energy scanning protocol apply to arterial phase scan of liver triple-phase examination, and improve the diagnosis rate of the lesion. The Toshiba CT has a 320-row detector, the lap maximum scan range of up to 16 cm, can be an extension of the wide volume scan. The study use Siemens CT and Toshiba CT to evaluate the effective dose of the liver triple-phase examination. In image quality evaluation, the study use the dual energy mode of Siemens CT and the wide-volume mode of Toshiba CT to scan Catphan phantom and reconstruct image by FBP and IR mode to compare their traditional spiral scan mode. The Rando phantom filled with TLDs to evaluate dose in the liver triple-phase examination. In addition, the experiment also used a single topogram scan protocol to assess the doses of the liver triple-phase examination. The Catphan phantom with self-made body annulus phantom was scanned by the AEC system and the 12 manual dose modes to evaluate the image quality of the four scan mode. In clinical examinations of the liver triple-phase, the lowest effective dose is in Siemens CT, the next is the wide-volume mode of Toshiba CT and the helical mode of Toshiba CT. In the FOMCNR of AEC mode reconstructed by FBP method, the best of FOMCNR is the dual energy scan mode of Siemens CT, and the next is the helical mode of Toshiba CT. If we use the IR mode of image reconstructions, the best of FOMCNR is also the dual energy scan mode of Siemens CT, and the next is the single energy scan mode of Siemens CT and the wide-volume mode of Toshiba CT. 陳拓榮 2012 學位論文 ; thesis 111 zh-TW
collection NDLTD
language zh-TW
format Others
sources NDLTD
description 碩士 === 中山醫學大學 === 生物醫學科學學系碩士班 === 100 === The traditional spiral scans of computer tomography using a single energy, and recently the Siemens CT using dual-energy scanning protocol apply to arterial phase scan of liver triple-phase examination, and improve the diagnosis rate of the lesion. The Toshiba CT has a 320-row detector, the lap maximum scan range of up to 16 cm, can be an extension of the wide volume scan. The study use Siemens CT and Toshiba CT to evaluate the effective dose of the liver triple-phase examination. In image quality evaluation, the study use the dual energy mode of Siemens CT and the wide-volume mode of Toshiba CT to scan Catphan phantom and reconstruct image by FBP and IR mode to compare their traditional spiral scan mode. The Rando phantom filled with TLDs to evaluate dose in the liver triple-phase examination. In addition, the experiment also used a single topogram scan protocol to assess the doses of the liver triple-phase examination. The Catphan phantom with self-made body annulus phantom was scanned by the AEC system and the 12 manual dose modes to evaluate the image quality of the four scan mode. In clinical examinations of the liver triple-phase, the lowest effective dose is in Siemens CT, the next is the wide-volume mode of Toshiba CT and the helical mode of Toshiba CT. In the FOMCNR of AEC mode reconstructed by FBP method, the best of FOMCNR is the dual energy scan mode of Siemens CT, and the next is the helical mode of Toshiba CT. If we use the IR mode of image reconstructions, the best of FOMCNR is also the dual energy scan mode of Siemens CT, and the next is the single energy scan mode of Siemens CT and the wide-volume mode of Toshiba CT.
author2 陳拓榮
author_facet 陳拓榮
Po-Hung
賴柏宏
author Po-Hung
賴柏宏
spellingShingle Po-Hung
賴柏宏
Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study
author_sort Po-Hung
title Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study
title_short Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study
title_full Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study
title_fullStr Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study
title_full_unstemmed Investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study
title_sort investigate liver triple-phase of the dual-energy and the 320-slice computed tomography image quality and radiation dose: the phantom study
publishDate 2012
url http://ndltd.ncl.edu.tw/handle/93808531008655773890
work_keys_str_mv AT pohung investigatelivertriplephaseofthedualenergyandthe320slicecomputedtomographyimagequalityandradiationdosethephantomstudy
AT làibǎihóng investigatelivertriplephaseofthedualenergyandthe320slicecomputedtomographyimagequalityandradiationdosethephantomstudy
AT pohung tàntǎoshuāngyuányǔ320qièdiànnǎoduàncénggānsānxiāngzhīyǐngxiàngpǐnzhìyǔfúshèjìliàngjiǎtǐshíyàn
AT làibǎihóng tàntǎoshuāngyuányǔ320qièdiànnǎoduàncénggānsānxiāngzhīyǐngxiàngpǐnzhìyǔfúshèjìliàngjiǎtǐshíyàn
_version_ 1718069862244286464