A Bf3 Fast Neutron Dosimeter (Long Counter)

<p>A BF3 fast neutron dosimeter or long counter is constructed and calibrated for use with the Cockcroft-Walton accelerator and UTR-IO reactor at Virginia Polytechnic Institute. The counter consists of a paraffin cylinder 8 inches in diameter, having a BF3 proportional detector on the axis and...

Full description

Bibliographic Details
Main Author: Leyland, Walter A.
Other Authors: Physics
Format: Others
Language:en
Published: Virginia Tech 2014
Subjects:
Online Access:http://hdl.handle.net/10919/42283
http://scholar.lib.vt.edu/theses/available/etd-04262010-020346/
Description
Summary:<p>A BF3 fast neutron dosimeter or long counter is constructed and calibrated for use with the Cockcroft-Walton accelerator and UTR-IO reactor at Virginia Polytechnic Institute. The counter consists of a paraffin cylinder 8 inches in diameter, having a BF3 proportional detector on the axis and a set of holes around the detector to improve the low energy response. External shields of paraffin, 3 inches thick, and boron-trioxide, 1/2 inch thick, enclose the paraffin cylinder. These shields serve to reduce the background effect and improve the high energy response.</p> <p> Calibration of the counter is accomplished using as neutron sources: plutonium-berylliwn and the induced reactions H2(d,n)He3 and H3(d,n)He4. Long counter efficiencies for both a 1 inch and 1/2 inch BF) tube yield a flat response over the energy range, 2.5 to 4 Mev whereas an approximately 60% decrease in efficiency is observed for a 14.1 Mev neutron.</p> <p> Count rate of long counter is expressed as a function of the incident neutron flux. Neutron fluxes, ranging from 0 to 1100 neutrons/cm2-seeond are monitored with the counter. </p> === Master of Science