Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair

The ability to accurately and in a repeatable fashion, measure aortic valve dimensions during aortic valve repair is critical to the restoration of function in a diseased aortic valve, as for example in aortic insufficiency. Although several methods for measuring aortic valve dimensions have been sh...

Full description

Bibliographic Details
Main Author: Qureshi, Rohail
Other Authors: Labrosse, Michel
Format: Others
Language:en
Published: Université d'Ottawa / University of Ottawa 2021
Subjects:
Online Access:http://hdl.handle.net/10393/42857
http://dx.doi.org/10.20381/ruor-27074
id ndltd-uottawa.ca-oai-ruor.uottawa.ca-10393-42857
record_format oai_dc
spelling ndltd-uottawa.ca-oai-ruor.uottawa.ca-10393-428572021-10-29T05:28:41Z Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair Qureshi, Rohail Labrosse, Michel Aortic valve repair Aortic valve dimensions Measurement device for aortic valve The ability to accurately and in a repeatable fashion, measure aortic valve dimensions during aortic valve repair is critical to the restoration of function in a diseased aortic valve, as for example in aortic insufficiency. Although several methods for measuring aortic valve dimensions have been shown to be feasible, they are approximate and lack the accuracy, robustness, and repeatability one would expect to support aortic valve repair surgery. In addition, they do not allow for the intra-operative measurement of aortic valves under conditions equivalent to the physiological (pressurized) state. A prototype medical device was designed, and 3-D printed at the University of Ottawa that would allow cardiac surgeons performing aortic valve repair to do just that. The prototype was tested for its accuracy and precision at the University of Ottawa Heart Institute using porcine aortic valves. Based on unsatisfactory results of this experimentation, namely, that the device was applying forces that were too large, a numerical simulation study was designed using a commercial finite element software LS-DYNA. This simulation study was used to explore the forces that the prototype device needed to apply to obtain end-diastolic pressurized dimensions of the aortic valve. The simulation study showed that one single device was likely not possible to obtain measurements in an aortic valve. However, a system of two devices could be imagined, one to measure the STJ diameter and free margin length of the aortic valve cusps, and one to measure the cusp height of the same valve, for the purposes of aortic valve repair. 2021-10-27T20:58:52Z 2021-10-27T20:58:52Z 2021-10-27 Thesis http://hdl.handle.net/10393/42857 http://dx.doi.org/10.20381/ruor-27074 en application/pdf Université d'Ottawa / University of Ottawa
collection NDLTD
language en
format Others
sources NDLTD
topic Aortic valve repair
Aortic valve dimensions
Measurement device for aortic valve
spellingShingle Aortic valve repair
Aortic valve dimensions
Measurement device for aortic valve
Qureshi, Rohail
Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair
description The ability to accurately and in a repeatable fashion, measure aortic valve dimensions during aortic valve repair is critical to the restoration of function in a diseased aortic valve, as for example in aortic insufficiency. Although several methods for measuring aortic valve dimensions have been shown to be feasible, they are approximate and lack the accuracy, robustness, and repeatability one would expect to support aortic valve repair surgery. In addition, they do not allow for the intra-operative measurement of aortic valves under conditions equivalent to the physiological (pressurized) state. A prototype medical device was designed, and 3-D printed at the University of Ottawa that would allow cardiac surgeons performing aortic valve repair to do just that. The prototype was tested for its accuracy and precision at the University of Ottawa Heart Institute using porcine aortic valves. Based on unsatisfactory results of this experimentation, namely, that the device was applying forces that were too large, a numerical simulation study was designed using a commercial finite element software LS-DYNA. This simulation study was used to explore the forces that the prototype device needed to apply to obtain end-diastolic pressurized dimensions of the aortic valve. The simulation study showed that one single device was likely not possible to obtain measurements in an aortic valve. However, a system of two devices could be imagined, one to measure the STJ diameter and free margin length of the aortic valve cusps, and one to measure the cusp height of the same valve, for the purposes of aortic valve repair.
author2 Labrosse, Michel
author_facet Labrosse, Michel
Qureshi, Rohail
author Qureshi, Rohail
author_sort Qureshi, Rohail
title Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair
title_short Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair
title_full Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair
title_fullStr Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair
title_full_unstemmed Evaluating the Feasibility and Effectiveness of a Measurement Device to Be Used Intraoperatively During Aortic Valve Repair
title_sort evaluating the feasibility and effectiveness of a measurement device to be used intraoperatively during aortic valve repair
publisher Université d'Ottawa / University of Ottawa
publishDate 2021
url http://hdl.handle.net/10393/42857
http://dx.doi.org/10.20381/ruor-27074
work_keys_str_mv AT qureshirohail evaluatingthefeasibilityandeffectivenessofameasurementdevicetobeusedintraoperativelyduringaorticvalverepair
_version_ 1719491753412657152