Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers

This study evaluated and compared the overall equipment effectiveness (OEE), sensitivity, specificity, and efficiency of the high-end hematology analyzers, Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10). A total of 400 anonymized left over's K2 EDTA whole blood samples were analyzed for co...

Full description

Bibliographic Details
Main Authors: Chandran Nair, S. (Author), Mammen, J.J (Author), Mullai, S. (Author), Murugan, P. (Author), Rastogi, S. (Author), Sukumar, A.S (Author)
Format: Article
Language:English
Published: Elsevier B.V. 2022
Subjects:
Online Access:View Fulltext in Publisher
LEADER 02414nam a2200265Ia 4500
001 10.1016-j.plabm.2022.e00275
008 220706s2022 CNT 000 0 und d
020 |a 23525517 (ISSN) 
245 1 0 |a Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers 
260 0 |b Elsevier B.V.  |c 2022 
856 |z View Fulltext in Publisher  |u https://doi.org/10.1016/j.plabm.2022.e00275 
520 3 |a This study evaluated and compared the overall equipment effectiveness (OEE), sensitivity, specificity, and efficiency of the high-end hematology analyzers, Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10). A total of 400 anonymized left over's K2 EDTA whole blood samples were analyzed for complete blood count. Of 400 samples, 200 were tested on Yumizen H2500; DxH 800 & DxH 900 while the other 200 were tested on Yumizen H2500 & XN-9000 (XN-10), respectively. The OEE was good and comparable for all the hematology analyzers except DxH 800 showing an average status. The sensitivity (%), specificity (%) and turnaround time (in minutes) for Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10) were 91.67, 61.11 & 103; 66.67, 54.55, & 149; 83.33, 27.27 & 136; 83.33, 28.57 & 122, respectively. Confusion matrix highlights the difficulty for DxH 800 and DxH 900 to discriminate left shift or blasts with large hyper-segmented neutrophils. The flags triggered by Yumizen H2500 were markedly changed to large hyper-segmented neutrophils. Lymphoblast caused more confusion for XN-9000 (XN-10), as it came out to be atypical lymphocytes, or hypersegmented neutrophils. Although comparable in OEE index to other analyzers, the Yumizen H2500 seems to be more reliable in detecting the abnormal cells as it has high sensitivity, specificity and less turnaround time. Thus, analysis adding specificity, sensitivity, and efficiency parameters to the OEE index provides more reliable information of the analyzers. © 2022 The Authors 
650 0 4 |a Efficiency 
650 0 4 |a High-end hematology analyzers 
650 0 4 |a Overall equipment effectiveness 
650 0 4 |a Sensitivity 
650 0 4 |a Specificity 
650 0 4 |a Yumizen H2500 
700 1 0 |a Chandran Nair, S.  |e author 
700 1 0 |a Mammen, J.J.  |e author 
700 1 0 |a Mullai, S.  |e author 
700 1 0 |a Murugan, P.  |e author 
700 1 0 |a Rastogi, S.  |e author 
700 1 0 |a Sukumar, A.S.  |e author 
773 |t Practical Laboratory Medicine