Overall equipment effectiveness, efficiency and slide review analysis of high-end hematology analyzers
This study evaluated and compared the overall equipment effectiveness (OEE), sensitivity, specificity, and efficiency of the high-end hematology analyzers, Yumizen H2500, DxH 800, DxH 900 and XN-9000 (XN-10). A total of 400 anonymized left over's K2 EDTA whole blood samples were analyzed for co...