Correction for a measurement artifact of the Multi-Angle Absorption Photometer (MAAP) at high black carbon mass concentration levels

The Multi-Angle Absorption Photometer (MAAP) is a widely-used instrument for aerosol black carbon (BC) measurements. In this paper, we show correction methods for an artifact found to affect the instrument accuracy in environments characterized by high black carbon concentrations. The artifact occur...

Full description

Bibliographic Details
Main Authors: A.-P. Hyvärinen, V. Vakkari, L. Laakso, R. K. Hooda, V. P. Sharma, T. S. Panwar, J. P. Beukes, P. G. van Zyl, M. Josipovic, R. M. Garland, M. O. Andreae, U. Pöschl, A. Petzold
Format: Article
Language:English
Published: Copernicus Publications 2013-01-01
Series:Atmospheric Measurement Techniques