Late Right Coronary Obstruction Following TAVR in a Degenerated Surgical Aortic Bioprosthetic Valve

We present a case of an 80-year-old woman with severe aortic insufficiency due to a degenerated 21-mm biological prosthetic aortic valve implanted 9 years earlier, treated by using a transcatheter aortic valve replacement valve-in-valve procedure and who experienced a delayed right coronary obstruct...

Full description

Bibliographic Details
Main Authors: Frederic Bouisset, MD, Bertrand Marcheix, MD, PhD, Etienne Grunenwald, MD, Thibault Lhermusier, MD, PhD
Format: Article
Language:English
Published: Elsevier 2019-10-01
Series:JACC: Case Reports
Subjects:
Online Access:http://www.sciencedirect.com/science/article/pii/S2666084919302505