Faulty nitrous oxide–oxygen interlock or hypoxic device leading to higher concentration of oxygen delivery

A “pre-use check” to ensure the correct functioning of anesthetic equipment is essential to patient safety. The anesthesiologist has a primary responsibility of checking the machine and equipment before administration of anesthesia. We report a case where pre-use check would have failed to check del...