RP-HPLC Method for the Estimation of Nebivolol in Tablet Dosage Form

A reverse phase HPLC method is described for the determination of nebivolol in tablet dosage form. Chromatography was carried on a Hypersil ODS C18 column using a mixture of methanol and water (80:20 v/v) as the mobile phase at a flow rate of 1.0 mL/min with detection at 282 nm. Chlorzoxazone was us...