Effect of Low Refocusing Angle in T1-Weighted Spin Echo and Fast Spin Echo MRI on Low-Contrast Detectability: A Comparative Phantom Study at 1.5 and 3 Tesla
MRI tissue contrast is not well preserved at high field. In this work, we used a phantom with known, intrinsic contrast (3.6%) for model tissue pairs to test the effects of low angle refocusing pulses and magnetization transfer from adjacent slices on intrinsic contrast at 1.5 and 3 Tesla. Only T1-w...