Transmission of Mycobacterium chimaera from Heater–Cooler Units during Cardiac Surgery despite an Ultraclean Air Ventilation System

Heater–cooler units (HCUs) were recently identified as a source of Mycobacterium chimaera causing surgical site infections. We investigated transmission of this bacterium from HCUs to the surgical field by using a thermic anemometer and particle counter, videotape of an operating room equipped with...

Full description

Bibliographic Details
Main Authors: Rami Sommerstein, Christian Rüegg, Philipp Kohler, Guido V. Bloemberg, Stefan P. Kuster, Hugo Sax
Format: Article
Language:English
Published: Centers for Disease Control and Prevention 2016-06-01
Series:Emerging Infectious Diseases
Subjects:
Online Access:https://wwwnc.cdc.gov/eid/article/22/6/16-0045_article