Selective and rapid determination of raltegravir in human plasma by liquid chromatographyâtandem mass spectrometry in the negative ionization mode

A selective and rapid high-performance liquid chromatographyâtandem mass spectrometry method was developed and validated for the quantification of raltegravir using raltegravir-d3 as an internal standard (IS). The analyte and IS were extracted with methylene chloride and n-hexane solvent mixture fro...

Full description

Bibliographic Details
Main Authors: Ajay Gupta, Swati Guttikar, Priyanka A. Shah, Gajendra Solanki, Pranav S. Shrivastav, Mallika Sanyal
Format: Article
Language:English
Published: Elsevier 2015-04-01
Series:Journal of Pharmaceutical Analysis
Online Access:http://www.sciencedirect.com/science/article/pii/S2095177914000872