A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature

We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors...

Full description

Bibliographic Details
Main Authors: David Morley, Ian Starks, Justin Lim
Format: Article
Language:English
Published: Hindawi Limited 2012-01-01
Series:Case Reports in Orthopedics