Upstream installation effects on metrological characteristics of flow sensors

The results of experimental investigation of upstream installation effects on metrological characteristics of the electromagnetic and turbine flow sensors of the nominal diameter DN50 are presented. The influence of inlet pipe diameter and influence of gasket diameter on accuracy of flow sensors we...

Full description

Bibliographic Details
Main Authors: Eugenijus Maslauskas, Gediminas Zygmantas
Format: Article
Language:English
Published: Vilnius Gediminas Technical University 2011-04-01
Series:Mokslas: Lietuvos Ateitis
Subjects: