Combined effect of dose gradient and rotational error on prescribed dose coverage for single isocenter multiple brain metastases in frameless stereotactic radiotherapy

Abstract Background To evaluate the combined effect of rotational error and dose gradient on target dose coverage in frameless stereotactic radiotherapy. Methods Three spherical targets of different diameters (1, 1.5, and 2 cm) were drawn and placed equidistantly on the same axial brain computed tom...